4 Maintenance
4
Maintenance
Exchange Spray Can
Insert spray can
Open side door.
Push securing ring up
Engage spray can flange in plastic holder
Push securing ring down
Only use original SciCan spray cans!
Close side door.
Remove spray can
Open side door.
Push securing ring up
Pull spray can from plastic holder
Close side door.
Disposal of spray cans
Completely empty spray can. Follow local regulations for disposal.
In Europe, the waste code number according to the European waste
catalogue is 15 01 04 – metals, packaging, empty.
Exchange Absorption Pad
Open front door.
Pull out and remove absorption pad from the two holding angles.
Hang new absorption pad into the two holding angles.
Only use original SciCan absorption pads.
Close front door.
Set Maintenance Time
To change maintenance time, switch on power while button
The LED
(Stand by) then toggles at 0.3-second intervals.
Maintenance time is set as follows:
15 s
LED place 1 illuminate
20 s
LED place 2 illuminate
30 s
LED place 3 illuminate
40 s
LED place 1, 2 and 3 illuminates
The
button is used to change the setting. The most recently set value is automatically saved after 5 seconds.
After saving, the unit assumes normal operation (LED
Disposal
See note disposal of consumables.
Cleaning
The stainless steel surfaces can be cleaned and serviced by means of a micro-fibre cloth and a stainless steel care
product (e.g. ECOLAB Chromol). The operating foil keypad is cleaned with a damp cloth (soaked in water).
Exchanging Fuses
Remove power plug from the STATMATIC.
Remove fuse drawer from power module
(push both flaps of the fuse drawer inward).
Replace fuses.
Insert fuse drawer again.
Only use fuse type described in the "Technical data" sheet!
Page 9
is pressed.
illuminates).
STATMATIC 31
EN