5. Select CAL-SLP.
6. Measure the Standard 1 vial as follows:
a. Enter the concentration for the Standard 1 vial (ppb) from the
Certificate of Analysis for the parameter selected in step 4.
b. Clean the Standard 1 vial with a no-lint cloth.
c. Install the Standard 1 vial in the vial adapter.
d. Put the vial cover on the sample vial.
e. Push Enter to measure the Standard 1 vial.
7. Measure the Standard 2 vial as follows:
a. Enter the concentration for the Standard 2 vial (ppb) from the
Certificate of Analysis for the parameter selected in step 4.
b. Clean the Standard 2 vial with a no-lint cloth.
c. Install the Standard 2 vial in the vial adapter.
d. Put the vial cover on the sample vial.
e. Push Enter to measure the Standard 2 vial.
8. Push Enter to save the results.
8.4 Set to the factory default calibration
To remove a user-entered slope calibration from the instrument and
use the factory slope calibration, do the steps that follow:
1. Select CALIBRATION, then select the parameter for the
calibration.
2. Select DEFAULT, then push Enter to confirm.
Section 9 Data management
9.1 Show the measurements on the instrument
1. Select SYSTEM > Log.
2. Selete the month the measurements were done, then push Enter.
3. Push the DOWN and UP arrows to scroll through the
measurements.