Set the sample probe position
The lift mechanism for the sample probe can be set to automatically go to the correct position for
sampling.
1
Put a sample container on the module. If a small vial is used, refer to
clamp.
2
Push the
3
Push the
4
Push the
correct bottom position.
NOTE
This position is recognized by the external software.
The bottom position is saved until a new bottom position is set.
Flush the syringe
Flush the syringe when a new syringe is installed or when a new sample is used. To optimize the
performance of the syringe:
• Do not operate the syringe for more than a few cycles without liquid.
• Thoroughly flush the syringe with distilled or deionized water after use.
1
Push the
2
Put the container with the flushing liquid under the sample probe.
3
Push the
4
Push the
the drain.
NOTE
The system can also be flushed with the external software.
DOC026.98.80097, Edition 7
button to lower the sample probe to the correct position.
SET
button to raise the sample probe to the top position.
LIFT/LOWER
button to automatically move the sample probe down and stop at the
LIFT/LOWER
button to raise the sample probe.
LIFT/LOWER
button to lower the sample probe in the flushing liquid.
LIFT/LOWER
button. The syringe fills with the flushing liquid and pushes the liquid out to
CLEAN
HIAC 9703+ Liquid Particle Counter