Figure 23 Calibration setting interface
Figure 24 Calibrate interface
Under [Calibrate] interface, push the syringe once, the device will display "REPEAT", then push the syringe once
again. After twice correct continuous operation, the calibrating will be succeed, and the device will display "OK!".
Finally the interface will jump to the former interface before calibration (The former interface: If the device is
calibrated after measurement completed, it will return to [Settings] interface; if calibrated before measurement
completed, it will return to [Testing] interface.).
If the device displays "Error! Please repeat", it indicates something wrong with the operation, please repeat the
calibrating until succeeded. If the device displays "Select right volume", please confirm whether the volume of
syringe and calibration selection is accordant, then repeat the calibrating until succeeded. If you need to stop
calibrating, just press the"Confirm" key to exit to the former interface before calibration.
Under [Calibration setting] interface, select "Adjust" to enter [Adjusting] interface, as shown in Fig.25. Press "Up"
or "Down" key to change the value (long pressing is available), then press "Confirm" key to return to [Adjusting
confirm] interface, as shown in Fig.26. Selecting "Yes" will save adjusted value, selecting "No" will cancel the
setting, then the device will return to [Calibration setting] interface.