VlowZ

The Vlow Z position measures AC voltage with lower
impedance (500kΩ) than normal voltage
measurement, this function is used to avoid
measuring ghost voltages.
3.12.

Other measurements

∆REL (relative mode)
Measures the relative difference. The relative mode
is available for the VLowZ, V, mV, Ω, C, A, mA, and
µA functions.
Hz/% (frequency/duty cycle)
Selects measurement of the frequency or duty cycle.
In the AC voltage or current mode, press the Hz/% key repeatedly to display the
frequency or duty cycle mode or return to the normal mode.
Frequency measurement in Voltmeter and Ammeter operation is available up to 1kHz.
Max-Min
In measurement mode, press Max-Min repeatedly to display Max, Min, or Max-Min. A
long press on Max-Min restores the normal mode.
4.
Reference conditions
Temperature
Humidity
Freq. for AC signal
Pure AC signal
Electric field
Magnetic field

TECHNICAL SPECIFICATIONS

+23°C ± 3°C.
45% to 75% RH
Supply Full battery (no low bat signal display) or accu.
1.5 VAA Ni-MH
45-65 Hz
<1 V/m
<40A/m
36