Specifications/Classifications - DeVilbiss Healthcare VacuAide 7310 Serie Manual De Instrucciones

Ocultar thumbs Ver también para VacuAide 7310 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 13
speCiFiCAtiOns
EN

SPECIfICATIonS/ClASSIfICATIonS

Size ..................................................................7.25"H x 7.25"W x 6.75"D (18.4cm x 18.4cm x 17.1cm)
Weight ....................................................................................................................... 3.375 lbs. (1.53 kg)
Electrical Requirements ................................. 100-240 VAC 50/60 Hz 0.75 A max; 12 VDC, 33 W max
Vacuum Range ...............................................................................................................50 to 550 mmHg
Air Flow @ pump inlet: ................................................................................. 27 LPM (free flow) typical
(may be less when running from internal battery)
Collection Canister Capacity.................................................................................................. 725 ml (cc)
Environmental Conditions
Operating Temperature Range ........................................................................32°F (0°C) - 104°F (40°C)
Operating Relative Humidity ......................................................................................................... 0-95%
Operating Atmospheric Pressure .................................................10.2 Psi (70 kPA) - 15.4 Psi (106 kPA)
Storage & Transport Temperature Range ................................................... -40°F (-40°C) - 158°F (70°C)
Storage & Transport Relative Humidity ......................................................................................... 0-95%
Storage & Transport Atmospheric Pressure ...................................7.3 Psi (50 kPA) - 15.4 Psi (106 kPA)
Warranty
7310 Series ...................................Two-years limited, excluding internal battery and collection canister
Internal Battery ......................................................................................................................... 6 months
Approvals
IEC 601-1; IEC 68; CAN/CSA-C22.2 No. 601.1-M90; UL 2601-1, CE EN 60601-1-2, ISO10079-1:1999
Meets RTCA/DO-160E .................................................................... DO-160E - section 21 Category M
(for battery operation only; commercial aircraft, airborne equipment)
Equipment Classifications
With respect to protection from electric shock ......................................Class I and internally powered
Degree of protection against electric shock ......................................................... Type BF Applied Parts
Degree of protection against ingress of liquids .....................................IP12 and ordinary power supply
Mode of Operation ........................................... Intermittent Operation: 30 minutes on, 30 minutes off
Equipment not suitable for use in the presence of a flammable anaesthetic mixture with air or nitrous
oxide.
ISO Classification
7310 Series - Electrically powered medical suction equipment for field and transport use according
to ISO 10079-1 : 1999
High Flow/High Vacuum
A-7310
10