4.2 Maintenance plan

Please check the battery/power pack voltage on a weekly basis.
For the battery/power pack voltage curve, please refer to the HMS CS operating manual.
Every 2 years:
As proof of continuous compliance to "Basic Requirements" pursuant to Directive 93/42/EEC, the Mobil-
®
O-Graph
blood pressure monitor must undergo metrological checks every two years. In certain
countries, this requirement may be regulated by national laws or regulations.
We can offer you to take care of metrological checks and the full service.
This service comprises:
Metrological checks
Software updates (if available)
Functional check: Electronics, pump and pneumatic circuit
Apart from the metrological check, no further maintenance measures for electromagnetic compatibility are
necessary.
Power packs:
Power packs are liable to ageing. Power packs that are damaged, or with which a 24-hour measurement is no
longer possible, must be replaced immediately.
We recommend that you only use the power packs supplied by IEM, the capacity and quality of which has been
tested. It is important to make sure that the capacity of the power packs is above 1500 mAh.
For further information, please refer to the operating manual for the charger
Care and maintenance
101
EN