EVENT
The EVENT button can be used to trigger an additional measurement to record an event that
affects blood pressure. The patient should record the circumstances of the event, such as
taking medication, for example, in a personal measurement log so that he/she can discuss the
events with you.
WARNING
After an automatic measurement, allow at least 3 minutes to elapse to avoid prolonged
impairment of blood circulation before actively starting a measurement.
START
The START button is used to activate 24-hour measurement and to perform a measurement
outside the predetermined measurement cycle.
WARNING
The values of the first measurement should be checked by the doctor for plausibility so that
subsequent correct automatic measurements can be carried out and correct cuff positioning
is ensured. In the event of an error measurement, please follow the instructions in chapters
3.4 "Preparation for measurement" and 5 "Troubleshooting".
If the START button is pressed, the display shows the number of previously registered
measurements and the manual measurement is started. It differs from the automatic
measurement according to the measurement log by a gradual inflation of the cuff. Here, the
required pressure of the cuff is determined at which the systolic blood pressure value can be
measured. This maximum required inflation pressure is stored and is immediately "started up"
by direct inflation during the subsequent automatic measurements.
The patient can use the START button to initiate manual measurements in addition to the
predetermined measurement cycle.