Battery Level And Recharging - Compex One Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 116

Battery Level and Recharging

EN
Never recharge the stimulator without first disconnecting the stimulation cables.
Always use the charger provided by Compex to recharge the battery.
The device runs on a battery. The battery's performance depends on the programmes
and stimulation energy used. We strongly recommend you to fully charge the battery
before using it for the first time as this will improve its performance and life span.
If the device is not to be used for more than 3 months, ensure the battery is fully
charged. If the device is not to be used for more than 6 months, ensure the battery is
fully charged and remove it from the stimulator. Before removing the battery, switch off
the stimulator.
Battery level
The battery icon on the bottom left of the screen indicates the level of charge in
the battery. The battery icon flashes when the battery is totally empty. It is no longer
possible to use the device. Recharge it immediately.
Recharging
Remove all stimulation cables from the stimulator before recharging it. Plug the charger
into a socket and connect the stimulator by sliding the red cover to the right to free the
charger connector.
The charge menu illustrated below appears automatically.
The duration of the charging operation is shown on the screen (a complete charge may
take 10 hours). The battery icon is animated while the battery is recharging. When fully
charged, the icon will be full and the total time taken to recharge the battery will flash
on the screen. Simply disconnect the charger: the stimulator will turn off automatically.
22

Publicidad

Tabla de contenido
loading

Tabla de contenido