1.4 Service Life

WARNING!
Risk of injury
Maintenance MUST be performed only by qualified
personnel.
Improper assembly may cause injury or damage.
– Regular maintenance of patient lifts and accessories is
necessary to assure proper operation.
– DO NOT overtighten the mounting hardware. This
will damage the mounting bracket.
The stand assist lift has an expected lifetime of 8 years when used
in accordance with safety instructions, maintenance intervals and
correct use, stated in this manual. The effective product lifetime can
vary according to frequency and intensity of use.
Actuator Service Life
Number of lifts per day
1–2
3
4
5
6
7–9
10–13
14–27
8
Service life of actuator
(in years)
10
9
6
5
4
3
2
1