EN-16
ATTENTION:
Please notice that you must enter the program as described under point 4.1.2 to change the
adjustments of the points 4.1.3 - 4.1.8. After you have stored the settings you can change to
normal program mode again by switching off the centrifuge for a short while.
All changed settings must be confirmed by the key "Start"(9). The word "Store"(12) appears in the display "RPM |
RCF"(A-1) - Only then the pre-selections are valid!
4.1.3
Temperature indication
Proceed as described under point 4.1.2 to enter this program mode and then press the key "Accel/Decel" (5). In
the display "Acc/Dec" (A-2) the word "Service" appears. Now select the letter "C" with the adjusting knob (1). As
a result, in the display "RPM | RCF" (A-1), the words "CELSI/temp" appear. If you press the key "RPM | RCF" (4),
the word "CELSI" flashes and you can change the display into Fahrenheit "FAREN", with the adjusting knob (1)
(See figure 32).
After you have stored the settings (See 4.1.2) you change back to the normal program mode again by
switching off the centrifuge for a short while.
Frontier
Figure. 30
A-1
Figure. 31
TM
Centrifuge FC5515/FC5515R
7
6
9
12