Internal load
Test current
switchable load
Switchable load
Automatic power
ON
Trip Test of RCD (earth leakage circuit breaker)
Test current
switchable load
Single Pole Phase Test
Voltage range
Frequency range
Acoustic
indication
Indication
Continuity Test (Rx) / Diode Test
Range
Tolerance
Test current
Acoustic
indication
Diode test
Indication
Overvoltage
protection
Automatic power
ON
Phase Rotation Indication
Voltage range
Frequency range
Indication
Resistance measurement (Ω )
Low resistance indication "
Resistance range
on LCD
Resolution
Approximately 2.4 W at 690 V AC/DC
(2100-Alpha / 2100-Beta)
Approximately 3.5 W at 1000 VAC /
Approximately 5.4 W at 1200 VDC
(2100-Gamma)
(load buttons not activated)
≤350 mA AC/DC at 690 V AC/DC
(2100-Alpha / 2100-Beta)
≤300 mAAC at 1000 VAC /
≤350 mADC at 1200 VDC
(2100-Gamma)
(load buttons activated)
Approximately 240 W at 690 V AC/DC
(2100-Alpha / 2100-Beta)
Approximately 300 W at 1000 VAC /
Approximately 420 W at 1200 VDC
(2100-Gamma)
(load buttons activated)
>12 V (activated load buttons)
>24 V (w/o load buttons)
(2100-Alpha / 2100-Beta)
>6 V (2100-Gamma)
>30 mAAC at 230 VAC (load buttons
activated)
180...690 VAC against earth
(2100-Alpha / 2100-Beta)
100...1000 VAC against earth
(2100-Gamma)
40 Hz...70 Hz
yes
Red LED
0....500 kΩ
0% to +50%
<5 μA
yes
yes
Red LED
690 V AC/DC (2100-Alpha / 2100-
Beta)
1000 VAC/1200 VDC (2100-Gamma)
< 500 kΩ
340...440 VAC phase to phase
(2100-Alpha / 2100-Beta)
170...1000 VAC phase to phase
(2100-Gamma)
47...63 Hz (2100-Alpha / 2100-Beta)
40...70 Hz (2100-Gamma)
Red LED
" (2100-Gamma only)
0...1999 Ω
1 Ω
12