Testing Impedances
The Impedance button is used to verify electrical integrity. Lead impedance is measured and
displayed for each of the Stimulator's 16 contacts. Impedances over 4500 Ω are considered to be
resultant from open or unconnected wires and are displayed with an orange bar on the Impedance
screen (or an "X" on the Program screen).
To check the Lead impedances:
1
Click
while in the Configuration screen or the Program screen.
The impedances will be measured and displayed. To re-measure the impedances, click
.
Note: All measured impedances are saved into the Clinician Programmer database.
2
To print the impedance report, click
3
To exit, click
.
.
Configuring the Vercise DBS System
Vercise™ DBS Programming Manual
91053785-07
19 of 323