Vercise™ DBS Programming Manual
3
Set the stimulation off time using the
The range for stimulation off time is 1 second to 90 minutes.
4
If desired, check Set as Patient Defaults to save these settings to the Remote Control as
defaults.
Note: If the Set as Patient Defaults is checked, these program settings will be used for all
future programs created for the current patient.
5
Click
to save your changes or
Defining the Stimulation Ramp Up Time
The stimulation Ramp Up time is the amount of time the Stimulator will take to increase stimulation
from 0 mA to the programmed amplitude when stimulation is turned on.
To define the Ramp Up time:
1
Check the box next to the Ramp Up option.
2
Set the Ramp Up time using the
The default setting is 3 seconds. The range for Ramp Up time is 1 to 10 seconds.
3
If desired, check Set as Patient Defaults to save settings to the Remote Control as defaults.
Note: If the Set as Patient Defaults is checked, these program settings will be used for all
future programs created for the current patient.
4
Click
to save your changes or
Activating or Deactivating Patient Amplitude Control
The Lockout Remote Amplitude option controls the patient's ability to change stimulation amplitude
using the Remote Control.
Note: The patient will be able to adjust their own amplitude level using the Remote Control
if the Lockout Remote Amplitude option is unchecked.
To activate or deactivate the Lockout Remote Amplitude option:
1
Check or uncheck the box next to the Lockout Remote Amplitude option.
Vercise™ DBS Programming Manual
91053785-07
38 of 323
buttons next to the Off field.
to exit without saving.
buttons next to the Ramp Up field.
to exit without saving.