Boston Scientific Vercise DBS Instrucciones De Uso página 9

Ocultar thumbs Ver también para Vercise DBS:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

4
Position the patient within 18 inches (45 cm) of the Remote Control and orient the large
surface of the Remote Control parallel to large surface of the Stimulator. Proper orientation
will ensure a reliable communication link between the Clinician Programmer and the
Stimulator.
Note: A message will be displayed on the Clinician Programmer if the communication
range is not satisfactory.
5
Turn on the Clinician Programmer and log in.
6
Click on the Vercise Launcher icon on the desktop.
7
Select the Vercise DBS System icon
Note: The Vercise Navigator may also be accessed on the same CP using Vercise
Launcher. Refer to the Vercise DBS Reference Guide for instructions on
which software application to use with your Deep Brain Stimulation System.
Refer to Vercise Navigator Manual with Vercise Navigator for instructions on
Vercise Navigator.
Note: Bionic Navigator and Vercise Navigator should not be run simultaneously on the
same CP.
8
Press and hold the
Note: The Remote Control should be linked to the Stimulator (see Linking a Remote
Control to a Stimulator on page 4).
9
Enter the Clinician's screen on the Remote Control by pressing and holding both the and
P
buttons simultaneously.
P
10
Press the
button to select "CP Mode."
The Remote Control will inform you that it is ready to communicate with the Clinician
Programmer.
Once the Remote Control is within range of the Stimulator, the Clinician Programmer will detect the
Stimulator and begin communication.
to start Bionic Navigator.
P
button to unlock the Remote Control.
Setting Up
Vercise™ DBS Programming Manual
91053785-07
3 of 323