8. Troubleshooting

Alpha Active 4
The following table provides a troubleshooting guide for the
system
in the event of malfunction.
Indicator
Possible Cause
Remedy
LOW PRESSURE
The tubeset is not connected
Check the tubeset connector
properly.
and ensure it is securely fitted to
the pump.
CPR not fully closed.
Close CPR unit.
There is a leak in the system.
Call service engineer.

POWER FAILURE

Power has been removed from
Re-apply power or switch the
the pump.
pump off.
SERVICE
Pump has detected an internal
Switch the pump off and call
fault.
service engineer.