Guidance And Manufacturer's Declaration - Electromagnetic Emissions - GETINGE GROUP ARJOHUNTLEIGH Alpha Active 4 Instrucciones De Uso

Ocultar thumbs Ver también para ARJOHUNTLEIGH Alpha Active 4:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Guidance and manufacturer's declaration - electromagnetic emissions

The pump is intended for use in the electromagnetic environment specified below. The customer or
the user of the pump should assure that it is used in such an environment.
Emissions Test
RF emissions
CISPR - 11
RF emissions
CISPR - 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Recommended separation distances between portable
and mobile RF communications equipment and the pump
The pump is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the pump can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the pump as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Compliance
Group 1
Class B
Class A
Complies
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2√P
0.12
0.38
1.2
2.8
12
Electromagnetic environment - guidance
The pump uses RF energy only for its internal
function. therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
The pump is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
m
80 MHz to 800 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
21
800 MHz to 2.5 GHz
d = 2.3√P
0.23
0.73
2.3
7.3
23

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

648311uk648319au

Tabla de contenido