Greenlee DML-54 Manual De Instrucciones página 17

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Accuracy
See the Specifications section for operating conditions and temperature coefficient.
Accuracy is specified as follows: ± (a percentage of the reading + a fixed amount)
at 23° C ± 5° C (73.4 °F ±9 °F), 0% to 75% relative humidity.
True RMS readings: Accuracies are specified from 5% to 100% of the range unless otherwise
specified. Frequency must be within the specified bandwidth for non-sinusoidal waveforms.
Crest factors are as follows:
• Crest factor < 3:1 at full scale
• Crest factor < 6:1 at half scale
AC Current
Range
499.9 µA
± (0.6% + 0.3 µA)
4999 µA
± (0.6% + 3 µA)
± (0.6% + 0.03 mA)
49.99 mA
± (1.0% + 0.3 mA)
499.9 mA
± (0.6% + 0.003 A)
4.999 A
± (0.6% + 0.03 A)
10.00 A*
* 10 A continuous; 20 A maximum (Duty Cycle: 30 seconds on, 5 minutes off)
DC Current
Range
499.9 µA
± (0.2% + 0.4 µA)
4999 µA
± (0.2% + 4 µA)
± (0.2% + 0.04 mA)
49.99 mA
± (0.2% + 0.4 mA)
499.9 mA
± (0.2% + 0.004 A)
4.999 A
± (0.2% + 0.04 A)
10.00 A*
* 10 A continuous; 20 A maximum (Duty Cycle: 30 seconds on, 5 minutes off)
Accuracy at
50 Hz - 60 Hz
Accuracy
Accuracy at
40 Hz - 1 kHz
± (0.8% + 0.4 µA)
± (0.8% + 4 µA)
± (0.8% + 0.04 mA)
± (1.0% + 0.4 mA)
± (0.8% +0.004 A)
± (0.8% +0.04 A)
Burden Voltage
0.15 mV/µA
0.15 mV/µA
3.3 mV/mA
3.3 mV/mA
0.03 V/A
0.03 V/A
DML-54 • DML-430
Burden
Voltage
0.15 mV/µA
0.15 mV/µA
3.3 mV/mA
3.3 mV/mA
0.03 V/A
0.03 V/A
17

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Dml-430

Tabla de contenido