Accuracy (cont'd)
Resistance
Range
49.99 Ω
499.9 Ω
4.999 kΩ
49.99 kΩ
499.9 kΩ
4.999 MΩ
50.00 MΩ
Capacitance
Range
49.99 nF
499.9 nF
4.999 µF
Notes: Accuracies are for film capacitors (capacitors with negligible dielectric absorption).
Measurements of larger capacitors can take up to 30 seconds.
Temperature (DM-820 only)
Range: –50 °C to 1000 °C
Accuracy: ± (0.3% + 3 °C)
Note: The accuracy information is for the meter only; see the information sheet provided with
your temperature probe (purchased separately) for its accuracy.
Diode Test
Measuring Range: 2.000 V
Test Current (Typical): 0.8 mA
Open Circuit Voltage: Less than 3.5 VDC
Accuracy: ± (1% + .001 V)
Continuity
Tone Threshold: Between 20 Ω and 200 Ω.
Response Time: Less than 100 µs
Crest Capture (Voltage and Current) for Crests ≥ 5 ms in duration
Accuracy: Specified accuracy ± 150 digits
Accuracy
± (0.2% + 0.06 Ω)
± (0.1% + 0.3 Ω)
± (0.1% + 0.002 kΩ)
± (0.1% + 0.02 kΩ)
± (0.1% + 0.2 kΩ)
± (0.4% + 0.003 MΩ)
± (1.5% + 0.05 MΩ)
Accuracy
± (0.8% + 0.03 nF)
± (0.8% + 0.3 nF)
± (1.0% + 0.003 µF)
DM-800 • DM-810 • DM-820
Open Circuit Voltage
< 3 VDC
< 1.3 VDC
Range
49.99 µF
± (2.0% + 0.03 µF)
499.9 µF
± (3.5% + 0.5 µF)
9999 µF
± (5.0% + 5 µF)
Accuracy
19