Operating Instructions CorroScout
1.5 Measuring the functionality of polarity protection diodes
1.6 Measuring the internal resistance of control sensors
of boilers and storage tanks in the range up a max. of 20 k .
Advice: When using a CORREX
where the titanium anode is always electrically insulated, the testing
of the applied drive potential as well as the determination of the
actually supplied protective current is directly possible with
CorroScout
When using Mg sacrificial anodes it is necessary that the sacrificial
anode is installed insulated. For carrying out the protective current
measurement in this case, the cable connection accessible from
outside between the Mg anode and the enamelled tank structure
should be opened for a short time so that the measuring device can
be connected in series with the circuit.
Sacrificial anodes welded into the lockscrew are not electrically
insulated and cannot be tested from outside with CorroScout
by means of electrical measurements. Here, a visual inspection is
necessary.

1.7 Temperature measurement especially for the temperature of water

between 0 °C and 100 °C.

2. Safety instructions

2.1
Measurements with CorroScout
specialists such as electricians, heating and sanitary plumbers!
2.2 Ensure before every voltage measurement, that the measuring
instrument is not in the current measuring range!
2.3 Always remove the test probes from the test object before
you are changing the measuring range.
22
®
500
®
500.
®
impressed current anode system
®
500 must be carried out by qualified
®
500