Operating Instructions CorroScout
After completed replacement, place the battery into the battery
chamber.
Carefully close CorroScout
of the connecting clip when closing the measuring instrument
Do not forget the rubber caps.
Put CorroScout
ning the housing safely with screws.
Dispose of used battery as special waste.
3.8 Replacement of the fuse
The following procedure is to be gone through when changing the
range fuse for the mA direct voltage input.
Disconnect CorroScout
switch it off.
Remove the measuring leads from the device.
Undo the fastening screws under the black rubber caps on the
back (bottom) of the device with a suitable screwdriver (Phillips type).
Carefully remove the lower shell of the housing locked at the top
of the device.
Remove the defective fuse and replace with a new one of the
same type and a nominal current strength of 0.315 A, quick-
acting, 250 V.
Carefully close CorroScout
of the connecting clip when closing the measuring instrument.
Do not forget the rubber caps.
Put CorroScout
ning the housing safely with screws.
3.9 "Hold" button
If the button "Hold" is actuated the actual measuring value will be
held. To quit this function, press the "Hold" button once again. The
device can then again be used for measurements.