QuICK rEFErENCE guIdE TrOuBLEShOOTINg
SYmpTOm
ACTION TAKEN
System will
1.
not operate:
No Power ON
2.
Indicator
3.
System will
1.
not operate:
Power ON
Indicator is
illuminated
2.
System operates:
1.
No water flow
2.
to insert tip
3.
or handpiece
4.
overheats
5.
System operates:
1.
No insert
2.
cavitation
3.
4.
5.
6.
Service
1.
indicator
blinking
2.
1.
System operates:
Service indicator
2.
illuminated
3.
1.
System operates:
2.
Purge mode will
not function --
icon flashing
Check that the Main Power Switch is in the ON (I) position, and that the detachable power cord
is fully seated in the receptacle on back of system.
Check that the system's power cord plug is fully seated in an appropriate AC wall outlet.
Check that the wall outlet is functional.
If the office has more than one foot control, test each to ensure that the proper foot control is
being used. With a handpiece and insert installed, depress the foot control to the first position.
The system should dispense water. If none of the foot controls operate the system, continue to
the next step.
Resynchronize one foot control to the system (see Directions for Use Section 6.8 Foot Control
Synchronization).
Assure that handpiece lavage control is properly adjusted.
Check for clogged insert; Replace insert if necessary.
Check that dental office water supply valves are open.
If the system is connected to DualSelect Dispensing System, check that fluid level in the selected
bottle is sufficient. Make sure valves are open when using external water source.
Check that the water line filter is clean; Replace filter, if needed.
Check that the Power Level Control is not in Rinse Mode.
Check the insert for damage and that it is properly installed in the Handpiece.
Check that the handpiece is properly installed to the cable assembly.
Verify that the handpiece's soft grip is flush with the hard plastic of the insert port. (Skip this step
if not using Soft Grip Accessory)
Turn the system's Main Power Switch OFF, (O) position. Wait 5 seconds and turn the system back ON.
If problem still exists, replace both "AA" batteries in foot control with new "AA" batteries, or
connect Auxiliary Foot Control Cable.
Slow blinking (1 blink per second) - The system is not operating within factory specifications.
A.
Remove insert.
B.
Turn Main Power Switch OFF, (O) position. Wait five seconds. Turn switch ON, (I) position.
C.
Operate Purge function.
D.
If service indicator still blinks, refer to Technical Support and Repairs to have unit serviced
as soon as possible.
Fast blinking (3 blinks per second) – Indicates improper set-up
A.
If insert is in the handpiece, remove. Verify the handpiece is properly seated and depress
the foot control for 2 seconds. If blinking stops, the system is ready for use. If blinking
remains, continue to the next step.
B.
Attach a NEW handpiece and depress foot control for 2 seconds. If blinking stops, the
system is ready for use. Discard the old handpiece or return if within warranty. If blinking
remains, continue to the next step.
C.
Install and fully seat an insert into handpiece. Depress foot control for 2 seconds. If
blinking stops, unit is ready for operation. If blinking remains, continue to the next step.
D.
IInstall and fully seat a NEW insert in handpiece and depress foot control for 2 seconds. If
blinking stops, system is ready for use. Discard old insert or return if within warranty.
If blinking remains, refer to Technical Support
Ensure that the base unit has adequate ventilation and is not near a heat source (i.e. radiator,
heat lamp, sunlight or other heat producing operatory equipment).
Turn Main Power Switch OFF, (O) position. Allow system to cool for 10 minutes and turn system
ON, (I) position. Verify light is not illuminated.
If light is still illuminated, refer to Technical Support and Repairs to have unit serviced as soon as
possible
Check that there is no insert in the handpiece.
Check that handpiece is properly installed to the cable assembly.