Operating Instructions Midwest
E Pro / Midwest
66 04 206 D3686.201.01.04.02
®
®
E Plus / Midwest
E mini
1. Brush the instrument under running water (< 38 °C, < 100 °F, at
least drinking water quality) and good lighting (min. 500 lux) and
color rendering index (min. 80 Ra) until no more dirt can be seen, for
at least 10 seconds.
2. Flush the drive channels with spray.
3. Maintain mechanical parts manually [ → 52].
4. Maintain the push button chuck manually [ → 54].
5. Pack the instrument in packaging material suitable for sterilization
and storage. e.g. paper/plastic composite packaging.
6. Conduct sterilization [ → 58].
7 Conditioning
51