7 Conditioning
56
Operating Instructions Midwest
7.4.3
Care of cooling spray nozzle openings
If your tap water is very hard, lime deposits may constrict or completely
block the cooling spray nozzles.
Interval
Interval
● At least once a week
1. Carefully clean the nozzle openings by running a cleaning wire
through them.
2. Let the instrument run briefly with cooling spray.
®
®
E Pro / Midwest
E Plus / Midwest
66 04 206 D3686.201.01.04.02
®
E mini