9.0
9.0 Specifi cations
Voltage range
Current input
Power cable
Power consumption
Operating modes
Stimulation time
Timer indication
Water temperature
Temperature indication
Feed temperature of the water
Water quality
Water fl ow in the irrigation mode
Water fl ow in the stimulation mode
Operating pressure
Connections
Operating time
Fuses
Protective earth conductor resistance
Earth leakage current
Enclosure leakage current
Patient leakage current
Ambient conditions
Transport/storage
Operation
Dimensions HxWxD
Weight
Protection class (EN 60601-1)
Degree of protection
Protection category
Classifi cation acc. to Annex IX
EEC directions 93/42/EEC
CE marking
Rules applied
UMDNS-Code
GMDN-Code
Reference-No.
UK
220 - 240 V~ ± 10 % 50/60 Hz
Special voltage: 100 - 127 V~ ± 10 % 50/60 Hz
6.8 A (220 - 240 V~); 14.8 A (100 - 127 V~)
3 m
1500 W
Thermal process for reducing bacterial count; Irrigation mode;
Stimulation mode; Heating off (energy saving)
adjustable by timer from 1 up to 99 sec.
Accuracy ± 0.5 s ± ½ digit
20°C - 47°C
Accuracy ± 0.5 °C ± ½ digit
+7...+19.5°C
Drinking-water
450 ml/min: standard unit and variant 250/450 ml/min
400 ml/min: variant 400/400 ml/min
500 ml/min: variant 500/500 ml/min
150 ml/min: standard unit 150/450 ml/min
250 ml/min: variant 250/450 ml/min
400 ml/min: variant 400/400 ml/min
500 ml/min: variant 500/500 ml/min
at least: +2000 hPa (2 bar); max: +6000 hPa (6 bar)
Power connection ; control output for a nystagmograph;
equipotential equalization; water supply G 3/4" external thread;
water drainage G 1/4" external thread; connection for double-barrelled hose
Continuous operation
quick-acting 8 A (220-240 V~); quick-acting 15 A (110-127 V~)
0,1 Ω max.
0.5 mA max.
0.1 mA max.
0.1 mA max.
-20...+50°C; 5...90 % humidity without condensing, air pressure 700...1060 hPa
+10...+35°C; 20...80 % humidity without condensing, air pressure 700...1060 hPa
14.5 x 37 x 31.5 cm
approx. 5.5 kg
I
Type B
IPX0
IIa
CE 0124
EN 60601-1: 1990 + A1:1993 + A2:1995
EN 60601-1-2: 1993 (EMV / EMC)
10-548
34891
502.0900.0
E
Issue of the Technical Specifi cations: 13.06.2000
Página 56
55