OHAUS Adventurer Serie Manual De Instrucciones página 9

Balanzas
Ocultar thumbs Ver también para Adventurer Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Error Codes
When a problem occurs using the balance, the display will indicated an error code. Review the listed codes and follow instructions to correct
the problem.
Err 2.0 Stability error. Check if balance is located near vibrating equipment or if air currents are affecting it.
Err 3.0 Calibration error (wrong mass). Incorrect or no calibration mass used when performing calibration procedure. Error will flash
momentarily, then balance will use previous calibration data. Recalibrate correctly.
Err 4
Invalid data checksum in type or adjust data. Return balance for servicing.
Err 6.0 Parts counting error - Average piece weight is less than 1d. Balance shows error then exits parts counting.
Err 8.4 Over or under load. Sample being weighed exceeds the capacity of the balance. If error occurs when the sample is within the
balance capacity, balance may be incorrectly calibrated. An underload such as the pan off of the balance could also display
Err 8.4. Recalibrate the balance.
Err 9
Internal data error. Return balance for servicing.
Err 9.8 Invalid checksum in calibration or setup data. The balance may need recalibration, particularly linearity calibration.
If the error persists after recalibration, thebalance must be serviced.
SPECIFICATIONS
Capacity
65g
Readability
Weighing units/modes
Repeatability (Std. Dev.)
Linearity
Tare Range
Stabilization Time (seconds)
Sensitivity Drift
Operating Temp Range
Power Requirements
Calibration
Display (in/cm)
Pan Size (in/cm)
Draft shield clearance height
Dimensions (WxHxD) (in/cm)
Net Wt (lb/kg)
Shipping Wt (lb/kg)
Item No.
AR0640
Notes 1. 65 g through 410 g balances include a three door draft shield.
2. Some models are not available in all countries.
Admissible ambient conditions
Temperature range:
Atmospheric humidity:
Altitude:
Voltage fluctuations:
Installation category:
Pollution degree:
Power supply voltage:
Power Adapter:
(1)
110g
(1)
210g
(1)
150g
(1)
210g
(1)
0.1mg
0.001g
g, kg, mg, ct, N, lb, oz, ozt, gn, dwt, momme, mesghal, 3 taels, ticals, Parts Counting
0.1mg
0.001g
±0.2mg
±0.3mg
±0.002g
4ppm/°C
3.5/9 dia.
9.3/23.6
above pan (in/cm)
8.5x14.3x13.5/
21.7x36.3x34.3
12.5/5.7
21/9.5
AR1140
AR2140
AR1530 AR2130 AR3130 AR4130
Use only in closed rooms
10 °C to 30 °C
Maximum relative humidity 80% for temperatures up to 31 °C decreasing linearly to 50%
relative humidity at 40 °C.
Up to 4000m
-15% +10%
II
2
8-14V ac, 50/60Hz, 6VA or 9.5-20V dc, 6W
AP3405
120V, 60Hz 10W
AP3405E
230V, 50Hz 80mA
AP3405B
240V, 50Hz 80mA
310g
(1)
410g
(1)
410/100g
(1)
510g
0.01/0.001g
0.01/0.001g
±0.01g/0.001g ±0.01g
Full capacity by subtraction
3 Seconds
10ppm/°C
50° to 86° F/10° to 30° C
8 - 14.5V 50/60Hz 6VA or 9.5 - 20V 6W
External Digital
LCD (1.2/3 high)
3.9/10 dia.
6.8/17.25
8.5x12.3x13.5/
21.7x31x34.3
10/4.5
16/7.3
ARRV70
AR5120 ARA520 ARB120 ARC120 ARRW60
5
1500g
2100
3100g 4100/1000g
4100g
0.01g
0.1/0.01g
0.1g
0.01g
0.1/0.01g
±0.02g
±0.1/0.01g
±0.1g
7.1/18 dia.
8.5x4.3x13.5/21.7x11x34.3
8.5/3.9
14/6.4
ARD110
0.1g

Hide quick links:

Publicidad

Tabla de contenido
loading