Connecting Power
NOTICE:
The socket/outlet must be installed near
the equipment and shall be easily accessible.
For use with an certified (CSA or equivalent) power
supply, which must have a limited and SELV circuit output.
Connect the AC Adapter cord to the connector located at the rear of the
balance and to a suitable power source.
Pan Installation
Place the pan support and pan on the balance. The balance is now
ready for operation. Balances with a draft shield do not have a pan
support.
Weigh Below Hook
For below balance weighing applications (eg. density determination),
a weigh below hook is installed at the bottom of the balance and is
part of the balance. To use, remove the protective plug cover located
at the bottom of the balance. The balance should then be supported
on an elevated level surface which allows access to the weigh below
hook. The balance should be releveled. Measurements are made by
the use of a fine wire attached to the internal weigh below hook.
CAUTION:
Do not attempt to remove the weigh below hook from the
balance as the balance may be damaged.
Security Bracket
A security bracket which is cast in the lower housing is provided at
the rear of the balance under the leveling bubble. This bracket allows
the balance to be secured by an optional cable and lock accessory.
OPERATION
Turning the Balance ON
Press >O/T<, , , , , all segments will appear briefly followed by a software
revision number (when plugged in first time) or after a power inter-
ruption and then * 0.00 g. Allow 20 minutes warm-up time. When
Not used
Stability Indicator
the balance is first turned on, it can be used to weigh in grams or
tare items without setting the menus.
Turning the Balance OFF
To turn the balance OFF, press and hold Mode Off until the display
indicates OFF then release.
Weighing
With the balance turned ON, it is ready to weigh in grams.
1. If other measuring units are desired, refer to Menu Section for setup
procedure.
2. If it is necessary to rezero the display, momentarily press >O/T<. . . . .
3. Place item(s) to be weighed on the pan and read the weight on
the display. The stability indicator * appears when the reading is
stable.
Taring
When weighing items that must be held in a container, taring sub-
tracts the container's weight from the total weight on the pan.
1. With an empty container on the pan, press >O/T< to zero the dis-
play.
2. As material is added to the container, the net weight is displayed.
Tared weight remains in balance memory until >O/T< is pressed
again.
MENU
The setup menu permits you to activate available weighing units,
parts counting, linearity calibration, keystroke tone and print param-
eters. To enter the menu the balance must first be turned off. Press
and hold the >O/T< until MENU is displayed then release. To
advance through the menus, press Mode Off. To enter a selected
menu, press >O/T<. After you make selections in any menu, you
may continue to make additional changes in other menus before
exiting and saving. To exit a menu and return to weigh mode without
saving settings, press and hold Mode Off until full display appears,
then release.
NOTE: Selecting END in the UNITS, SYS and PRINT menus will save
the settings for each menu.
MENU
UNITS - g, kg, mg, ct, N, lb, oz, ozt, GN, dwt, mo, m,
Hong Kong taels, Singapore taels, Taiwan taels, cL,
and PCS. - Set each to ON or OFF. NOTE: Only one Tael
may be on at a time. See specifications table for
available units.
LIN -
Performs linearity calibration.
SYS -
Permits setting tone ON or OFF, with each keystroke.
PRINT - Permits setting communication and print parameters.
MENU - Exit menu and return to weigh mode.
END
Activating Units and Parts Counting (PCS)
1. With the balance OFF, press and hold >O/T< until MENU is dis-
played, then release, UNITS
UNITS is displayed.
UNITS
UNITS
UNITS
2. Press >O/T<, On g is displayed. To turn this unit of measure ON or
OFF, press Mode Off to select ON or OFF condition.
3. To advance through all of the measuring units including parts count-
ing and to set each ON or OFF, press >O/T< to advance to next
unit then press Mode Off to select ON or OFF. . . . . You must continue
until END appears. NOTE: Only one Tael can be activated at a
time.
4. When END is displayed, press >O/T< to store unit selections.
5. To exit the menu, repeatedly press Mode Off until MENU END is
displayed, then press >O/T<. Balance returns to a weihging mode.