Installation

4.1

Purge air supply

The sensor needs a purge air supply for correct functioning.
The purge gas also keeps the measuring gas chamber dry
and prevents humidity build up that could harm the sensor.
Requirements of the purge gas supply:
Any air supply that fullfills the requirements below is suitable
to be used with this sensor:
– Delivered pressure: min. 1 – 2 barg
– Particulate free (40 micrometer)
– Condensate free (dew point below – 10 °C)
Connecting the purge air to the sensor
To connect the purge gas supply to the sensor
use a flexible nylon, PVC or a stainles steel hose.
The sensor is supplied with a 6 mm Swagelock connection
for this purpose.
4.2

Connecting the cable to the transmitter

CO
RS 485 cable
Color Function
brown 24 DC +
black
gray
yellow shield
blue
white RS 485 +
h
Notice: Cable assignement can be found in the
METTLER TOLEDO cable instruction manual.
h
Notice: For connecting the cable to the terminals of the
transmitter, please also refer to the instructions given in the
METTLER TOLEDO transmitter manual.
5

Operation

In case of any failure or malfunction, the equipment in
which the sensor is installed needs to be in a safe condition
(depressurized, empty, rinse, vent, etc.).
© 03 / 16 Mettler-Toledo GmbH
30 313 397
Transmitter
2
24 DC –
shield
RS 485 –
InPro 5500 i CO
VP or AK9 cable
Transmitter
M400
Channel 2 Channel 4
TB4
TB2
1
9
2
10
6
12
6
15
7
13
8
14
InPro 5500 i CO
Printed in Switzerland
Sensors
2
M800
TB4
9
10
12
15
13
14
Sensors
2

Publicidad

loading