For all ME EQUIPMENT and ME SYSTEMS
Radiated RF
3 V/m
IEC 61000-
80 MHz
4-3
to 2.5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
Table 3 - Continued
Recommended separation distance:
d=1.2 P
d=1.3 P
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
3 V/m
transmitter manufacturer and d is the
recommended separation distance in
meters (m). Field strengths from fixed
RF transmitters, as determined by an
electromagnetic site survey
be less than the compliance level
each frequency range. Interference
may occur in the vicinity of equipment
marked with the following symbol:
27
80 MHz to 800 MHz
800 MHz to 2.5 GHz
a
, should
b
in