For all ME EQUIPMENT and ME SYSTEMS
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in
the location in which the KD-795B is used exceeds the applicable RF compliance
level above, the KD-795B should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting
or relocating the KD-795B.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3 V/m.
Table 3 - Continued
28