Regulatory notes
Information about approvals can be found via this menu item.
Calibration
Set the date of the next calibration.
Factory settings
The device can be reset to the factory settings in this menu item.
Multi-Dispensing (MULTI-DISP)
For information on the function of this mode, see „Overview of modes"
→ 65.
For information on setting the volume, speed, and other operating
steps found in all modes, see „Operation" → 67.
Aspirating liquid
Condition
The STEP volume has been set via the
button.
a. Hold the tip vertically in the vessel.
During aspiration, make sure that the tip
opening is always covered by liquid, in
order to prevent air bubbles in the tip.
b. Press the STEP button.
The device aspirates the liquid until the
set volume or the nominal volume is
reached.
The
icon displays the number of
possible STEPs.
The
icon displays the available
volume.
Stopping and resuming liquid aspiration
Stopping liquid aspiration
a. To stop liquid aspiration, briefly press the STEP button.
Resuming liquid aspiration
a. Tap on the
button.
b. Press the STEP button 2 times in quick succession.
The device aspirates liquid.
75