Options
a. In the operating mode, tap on the symbol
The Options menu appears.
Option
'Manually/STEP Vol-
ume'
'Show last volumes'
'Help Work Mode'
You can save settings in Favorites, in order to open them again later, e.g.
aliquot lists, volume settings.
Cleaning
The device is not autoclavable.
The device is factory-calibrated and maintenance-free.
When the device is dirty, clean the outer surface with a moist cloth . Use
water or a mild detergent solution. Avoid acidic or aggressive cleaners.
Disassembly of the device is not permitted.
UV disinfection
The device is resistant to normal exposure to a UV disinfection lamp. The
device may change color due to the UV exposure.
UV disinfection recommendations:
Light spectrum
Wave length
Exposure time per cm
and intended inactivation rate
2
Meaning
Switch to titration mode
Displays the last volumes titrated.
Shows the help text for the mode.
Cleaning and disinfection
NOTICE
.
Favorites
UV-C
220 nm ... 270 nm
2 s ... 300 s