Shut down at the end of the day
1. Flush the
Dentsply
according the dental unit manufacturer's
instructions.
2. Remove any ultrasonic inserts and the Steri-Mate
handpiece. Clean and sterilize the handpiece and
all inserts used during the procedure (refer to
the Infection Control information booklet and the
Cavitron Ultrasonic Inserts Directions for Use for
cleaning and sterilizing instructions).
3. Clean and disinfect the handpiece cable assembly
by applying a medically approved non-immersion
type disinfectant solution,* carefully following the
instructions provided by the solution manufacturer.
To clean the cable, generously spray disinfectant
solution on a clean towel and wipe the cable and
connector. Discard used towel. To disinfect the
system, generously spray disinfectant on a clean
towel and wipe the cable and connector. Allow the
disinfectant solution to air dry.
4. Clean and disinfect all surfaces of the dental control
head according to the manufacturer's instructions.
5. Place the cable connector into the handpiece holder
for storage.
6. Turn off the power and water to the dental unit.
*NOTE: Water-based disinfection solutions are preferred.
Some alcohol-based disinfectant solutions may discolor the
handpiece cable and connector.
ENGLISH • 12
88016 Rev. 5 (0119).indd 12
Cavitron built-in ultrasonic system
4/11/2019 8:02:40 PM