• Incoming water temperature to the Cavitron system
should not exceed 25°C (77°F). If needed device
should be installed to maintain a temperature
within this specification, or a Cavitron DualSelect
Dispensing System attached to allow this system to
be operated as a closed water system.
• A manual shut-off valve on the dental water system
supply line should be used so that the water can be
completely shut-off when the office is unoccupied.
• A filter in the dental water system supply line
is recommended so that the particles in the
water supply will be trapped before reaching the
ultrasonic scaler.
• After the above installations are complete on
the dental water supply system, the dental office
water line should be thoroughly flushed prior to
connection to the ultrasonic system.
• After flushing system, verify there are no leaks.
Section 8: Ultrasonic Scaling
System Description

8.1 System Controls

The Dentsply Sirona Cavitron Ultrasonic Scaling System
module is enabled whenever the clinician removes the
handpiece from the handpiece holder.
88016 Rev. 5 (0119).indd 7
ON/OFF Function
The Ultrasonics is switched ON and OFF using the
dental unit foot control.
Power Adjustment
There are 2 options for adjusting the power to your
Cavitron Built-In Ultrasonic Scaling System.
Option 1: Power is adjusted by turning the power
control knob located on the dental unit control that
installed by your service technician.
Stickers have been provided for application
to the treatment unit to denote the
Cavitron Built-In Ultrasonic Scaling System
power range (see image). These stickers
are not required and may or may not have
been applied by your technician. If they have not been
applied or left behind by the technician and you would
like to apply them to your dental unit, contact Dentsply
Sirona Customer Service to order them.
Option 2: If you are using a treatment unit equipped
with a touch pad, your service technician will wire
the power adjustment to the appropriate buttons on
the dental unit. Power can be adjusted using these
buttons.
Lavage Flow Adjustment
Water flow through the handpiece is adjusted by
rotating a control element on the cable connector (blue).