Error message
Device disconnec­
ted
No new device
found
Cause
A current connec­
tion to the pros­
thesis was inter­
rupted.
No new prosthesis
could be found
since this pros­
thesis is already
stored.
Correction
3) See section "Managing pros­
theses" (see Page 35) for fur­
ther information.
Check the following points:
Distance from the prosthesis
to the remote control
Was the prosthesis connec­
ted to another remote con­
trol?
The
connection
already saved in the remote
control must be deleted in
order to re-connect the pros­
thesis to this remote control.
A new connection must then
be established from the pros­
thesis to this remote control
(see Page 35).
Use the
 key to confirm the
error message and try to
establish
again.
1) Confirm error message with
the  key.
2) Use the
,
the
menu
device" and confirm this with
the  key.
3) Use the
,
the prosthesis required and
confirm this with the  key.
that
was
the
connection
 keys to select
option
"Select
 keys to select
Ottobock | 43