8.2 Charger and Li-ion battery

Malfunction
Charger not working
Li-ion battery cannot be loaded
10
Detection
The "Battery" LED indicator is not
lit up
Li-ion battery not loaded
Li-ion battery plugged in, charge
state indicator does not light up
Cause
Mains power cord not plugged in
Mains power cord defective
Li-ion battery defect
Charger contacts dirty
Charger contacts damaged
Li-ion battery defect
Charger defect
Too high a temperature is measured
in the lithium ion battery during the
charging process
Battery charging error or Li-ion
battery defect
Remedy
Plug mains power cord into the
device socket on the charger and
into the socket of the mains power
supply
Replace mains power cord
Send Li-ion battery to
manufacturer for repair, see
Technical Service
Clean the contacts in the charging
bay, see Reprocessing
Send charger to manufacturer for
repair, see Technical Service
Send Li-ion battery to
manufacturer for repair, see
Technical Service
Send charger to manufacturer for
repair, see Technical Service
Remove the Li-ion battery from the
charging bay, leave to cool and
repeat the charging process
If the problem occurs again: Send
Li-ion battery to manufacturer for
repair, see Technical Service
Remove the Li-ion battery from the
charging bay and repeat the
charging process
If the problem occurs again: Send
Li-ion battery to manufacturer for
repair, see Technical Service