3.2.2
Chin support
Lock / release knob
for bite block
PaX-i3D Green (PHT-60CFO) User Manual
Completing Patient Positioning
1. After confirming patient positioning and beam alignment, adjust the temple
supports to fit snugly on both sides of the patient's head by clicking the temple
support adjustment button.
2. Click the READY button after the patient has been properly positioned. No
X-Ray will be emitted at this point.
3. Proceed to section 3.3 Initiating X-Ray Exposure.
TMJ Open
There are two TMJ sub modes: TMJ Open and TMJ Close. Scan a TMJ Open
image first and then scan a TMJ Close image. To prepare and position the
patient, follow the process outlined below.
For a child with a large head circumference, Man/Woman
mode may be selected instead of Child.
NOTE
Positioning the patient
Disinfect the Chinrest using an alcohol-based cleaning
solution and wipe away all residues with dry cloth before
CAUTION
proceeding any further.
3 Acquiring PANO Images
1. Insert the Chin support into the Chinrest
receptacle on the equipment.
2. Open the Temple Supports by clicking
the temple support adjustment button.