Inserting the Chemical Indicator
A Chemical Indicator must be used for each disinfection cycle and can only be used once.
After correctly loading the probe into the chamber, a chemical indicator shall be placed
into the holder. Refer to the Chemical Indicator Instructions For Use (IFU).
B2.5
Closing the Chamber Door
 The door will automatically lock at the start of a high-level disinfection cycle.
 If the door is not properly closed, screen message: CLOSE CHAMBER DOOR
B2.6
Disinfecting the Probe
Screen message: IS THE PROBE CLEAN AND DRY?
YES
1. If the probe has been pre-cleaned and dried according to section B2.2 above, press
YES using the soft key button. The device will then check device readiness to perform a
disinfection cycle. Once ready screen message: PRESS START TO BEGIN
2. Press the START button to initiate the cycle or CANCEL using the soft key buttons to
unlock the machine and remove the probe
3. The progress of the disinfection cycle is indicated on the screen
4. Screen message: DISINFECTING
5. High Level Disinfection cycle will take 7 minutes to complete.
NO
1. If the probe has NOT been pre-cleaned and dried according to section B2.2, select NO
using the soft key button
2. Screen message: REMOVE AND CLEAN THE PROBE
3. Remove the probe and complete pre-clean as directed in section B2.2. Then follow
instructions from section B2.3.
NOTE: If mist is visibly escaping from the chamber, remain at a distance from the device until
completion of the operating cycle, and until the mist stops. Do not come into direct contact
with the mist. Contact your customer service representative (see Part C - Troubleshooting).
B2.7
Removing the Probe
It is important to wear clean gloves when handling probes before and after a disinfection
cycle.
1. When the cycle has been successfully completed, the device will sound an audible
alarm. Screen message: CYCLE COMPLETE REMOVE AND WIPE PROBE
NOTE: If the probe is not removed within 1 minute of the cycle being complete, the warm-up
time will increase for the subsequent cycle.
2. Wear new gloves. Remove the used Chemical Indicator from the device and compare
colour change against the chart on the Chemical Indicator carton (refer to the
Chemical Indicator IFU for further instructions).
3. Discard used Chemical Indicator
4. Remove the probe carefully using minimal contact after the cycle is complete. Avoid
touching the probe against the chamber's hot surface
®
trophon
EPR User Manual – N00020
L00034 1.0 03/2014