Invacare® Perfecto₂™Series
2.3 Radio Frequency Interference
WARNING!
Risk of Injury or Damage
To reduce the risk of injury or product damage from
interference with wireless equipment:
– Keep concentrator at least 9.8 ft (3.0 m) away from
wireless communication equipment such as wireless
home network devices, mobile phones, cordless
phones and base stations, walkie-talkies, etc.
This equipment has been tested and found to comply with EMC limits
specified by IEC/EN 60601-1-2. These limits are designed to provide
a reasonable protection against electromagnetic interference in a
typical medical installation.
Other devices may experience interference from even the low levels
of electromagnetic emissions permitted by the above standards.
To determine if the emissions from the concentrator are causing
the interference, turn the concentrator Off. If the interference
with the other device(s) stops, then the concentrator is causing the
interference. In such rare cases, interference may be reduced or
corrected by one of the following measures:
•
Reposition, relocate, or increase the separation between the
equipment.
•
Connect the equipment into an outlet on a circuit different from
that to which the other device(s) is connected.
12
1163146-H~07