Invacare HomeFill II Oxygen System IOH200AW Manual Del Usuario página 23

Ocultar thumbs Ver también para HomeFill II Oxygen System IOH200AW:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 71
1.
Make sure the concentrator is On. Refer to the concentrator
User Manual.
The maximum oxygen concentrator flow rate for
patient use is listed in the table below. Otherwise, the
oxygen level to the compressor will be <90% and the
compressor will not fill the cylinder.
Invacare Oxygen
Concentrator
Platinum 5
Platinum 9
Perfecto
2
Patients with a higher prescribed flow should not refill a
cylinder while receiving supplemental oxygen treatment
from the oxygen concentrator.
2.
Set the concentrator flow rate to the patient to 2.5 L/min or less
for Platinum 5 and Perfecto
Refer to the concentrator User Manual.
3.
Make sure the cylinder is connected to the compressor. Refer
to Connecting the Cylinder to the Compressor in this section
of the manual.
Push the power switch A on the control panel to the On
4.
( I ) position.
1195663-A
Maximum Output Flow
for Patient Use
2.5 L/min
5 L/min
2.5 L/min
and 5 L/min or less for Platinum 9.
2
5.
Examine the control panel B. The following sequence of events
should occur:
0‐3 minutes ‐ The O
a.
2
be lit while the compressor warms up.
After three minutes ‐ The FILLING (GREEN) light D will be
b.
on while the cylinder is filling.
The FULL (GREEN) light E will be on when the cylinder is
c.
finished filling. Proceed to STEP 6.
The O
BELOW NORMAL (YELLOW) light may come
2
on if the oxygen level from the concentrator has not
yet reached or drops below 90%. The compressor will
begin or resume filling when the oxygen level reaches
90%. If light does not go Off within 10 minutes, refer to
Troubleshooting in this manual.
If the ATTENTION (RED) light and audible alarm F are
on refer to Troubleshooting in this manual.
6.
Remove the full cylinder. Refer to Disconnecting the cylinder
from the Compressor in this section of the manual.
7.
Push the power switch to the Off (O) position.
For cylinder operation and use, refer to the specific user
manual for the cylinder/regulator assembly.
Below Normal (YELLOW) light C will
Usage
23