2.4 Service life

Based on ISO 10328, the prosthetic components were tested by the manu­
facturer for 3 million load cycles. With mobility not involving sports activities,
this would correspond to a service life of 3 to 5 years in an everyday pros­
thesis. A general service life for this prosthetic component as part of a sport
prosthesis cannot be determined, since strain varies widely depending on
the sport and intensity of use.
3 Safety
3.1 Explanation of warning symbols
Warning regarding possible risks of accident or injury.
CAUTION
Warning regarding possible technical damage.
NOTICE
3.2 General safety instructions
CAUTION
Excessive strain on the product
Risk of injury due to breakage of load-bearing components
► Use the product according to the specified area of application (see
Page 16).
CAUTION
Unallowable combination of prosthetic components
Risk of injury due to breakage or deformation of the product
► Only combine the product with prosthetic components that are
approved for that purpose.
► Based on the instructions for use of the prosthetic components, verify
that they may be combined with each other.
CAUTION
Exceeding the service life
Risk of injury due to change in or loss of functionality and damage to the
product
► Ensure that the approved service life is not exceeded.