CORRECTING DATA ON A PROGRAM THAT IS RUNNING
7.11
If for some reason the operator decides to change the data once the program has already been
launched, he/she must proceed as follows:
A. Press START-STOP (20): the word OFF appears on the display (11).
B.
Press PROG. (19): the legend "Pro" appears on the display (11).
C.
Press ENTER and the following occurs on the control panel:
• the green horizontal segment of the recalled phase comes on.
• the first of the three parameters (i.e. temperature) (11) appears.
• the number of the recalled phase appears on the display (15).
• the green PROG. LED (27) flashes.
D. Change/cancel one or more of the phase data items.
E.
Save all changes by pressing ENTER.
F.
Press START/STOP (20) to return to the display (11): F appears.
Key in the number of the phase from which you wish to begin.
G. Press START/STOP (20).
CHANGING THE PROGRAM IN PROGRESS
7.12
If an undesired program is launched by mistake, proceed as follows:
A. Press START-STOP (20): OFF reappears on the display (11).
B.
Press SELECT./PROG. (12): "P" appears on the display (14).
C.
Press the correct program key: the corresponding number appears on the display (14).
D. Set the oven start time as described in Chapter 4 "Programming Oven Start Time".
CHECKS DURING WORK
7.13
Several checks can be run while the oven is working:
A. During phase 1 / Delayed start on
1. Pressing TEMP. (17) on the display (11) causes the current in-chamber temperature to appear.
• after a few seconds the display returns to its previous state.
168 of 272
MAINTENANCE AND USE MANUAL
EN
FURNACE DM