Installation; Operation; Standby Button - Sunoptic Surgical LED-4000 Manual Del Operador

Fuente de luz led
Ocultar thumbs Ver también para LED-4000:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 58

5. INSTALLATION

5.1 SETTING UP THE LED LIGHT SOURCE
Place the LED light source on a stable surface (cart, counter, stand, etc.).
NOTE
Avoid places where the light source may be splashed with liquid.
Absolutely DO NOT use in any environment with explosive or flammable
gases.
DO NOT block the venting grids of the light source or impede exhaust air flow. Users
must ensure that ventilation air temperatures are below the maximum rating for the
light source and are not influenced by warm exhaust from other sources.
Make sure the power switch is in OFF position.
Connect AC power cord to the power inlet located on the rear panel of the light source.
Use only cords provided with the light source.
CAUTION
Plug the supplied AC power cord into a wall outlet.
To prevent electric shock, connect power cords of peripheral equipment
through medical isolation transformers.
CAUTION
NOTE
When using medical isolation transformer, be sure to check the transformer power
ratings. Make sure that the power cord is connected to the main power a suitable
plug (USA use UL2601-1 rated isolation transformers and/or power strips only).
5.2 CONNECTING THE FIBER OPTIC CABLE
Connect the fiber optic cable to the instrument and then plug the light guide endfitting into the
appropriate turret port on the front panel.
NOTE
The turret has a light guide sensing feature that will not allow the LED light to turn on
until a fiberoptic cable is inserted into the active turret port.

6. OPERATION

6.1 POWERING UP THE LED LIGHT SOURCE
To operate the LED light source:
Turn on the power switch, the indicator light in the power switch will illuminate.
6.2 STAND-BY BUTTON
The STAND-BY button allows the user to turn the LED on or off when the main power switch is on,
provided a light guide is inserted in the active port.
Press AND HOLD this button momentarily for operation. An indicator lamp will be lit when stand-by
mode is active (LED off) and the Light Intensity Control and Display will remain on and adjustable.
LIT-216 Sunoptic Surgical
Rev. C
®
Page 10 of 114
(English)

Publicidad

Tabla de contenido
loading

Tabla de contenido