- Proximity Fields from RF Wireless Communications Equipment
Test
Band
frequency
(MHz)
(MHz)
385
380 - 390
450
430 - 470
710
745
704 - 787
780
810
870
800 - 960
930
1720
1845
1700 - 1990
1970
2450
2400 - 2570
5240
5500
5100 - 5800
5785
PC-BMG 3018_IM_new
Table 3
Immunity test levels
Professional healthcare
facility environment
Pulse modulation 18 Hz, 27 V/m
FM, ± 5 kHz deviation, 1 kHz sine,
28 V/m
Pulse modulation 217 Hz, 9 V/m
Pulse modulation 217 Hz, 28 V/m
Pulse modulation 217 Hz, 28 V/m
Pulse modulation 217 Hz, 28 V/m
Pulse modulation 217 Hz, 9 V/m