CAUTION: In the event of an alarm or you ob-
serve the Oxygen Concentrator is not working
properly; consult the troubleshooting section of
this manual. If you cannot resolve the problem,
consult your Equipment Provider.
CAUTION: If the humidifier bottle tubing is not
properly connected to the humidifier bottle
fitting or to the oxygen outlet, an oxygen leak
can occur.
CAUTION: Normally, you should not need to ad-
just the flowmeter on your unit. If you turn the
flowmeter adjustment knob clockwise, you will
decrease and can shut off the flow of oxygen
from your unit. For your convenience, the flow-
meter is marked in ½ LPM increments. For units
equipped with the 2 LPM flowmeter option, the
flowmeter is marked in 1/8 LPM increments for
flow settings up to 2 LPM.
Note: . Ensure the cannula is fully inserted and
secure. You should hear or feel oxygen flow to
the prongs of the nasal cannula. If oxygen does
not seem to flow, first verify that the flow meter
ball is registering a flow. Then, place the tip of the
cannula into a glass of water; if bubbles come out
of the cannula, oxygen is flowing. If bubbles do not
appear, refer to the troubleshooting section of this
manual.
Note: Always follow the cannula manufacturer's
instructions for proper use. Replace the disposable
cannula as recommended by the cannula manufac-
turer or your Equipment Provider. Additional sup-
plies are available from your Equipment Provider.
Note: Allow the unit to run for at least 5 minutes at
2 LPM or above before use.
Note: Always follow the cannula manufacturer's
instructions for proper use. Replace the disposable
cannula as recommended by the cannula manufac-
turer or your equipment provider. Additional sup-
plies are available from your equipment provider.
Note: The Manufacturer does not recommend the
sterilization of this equipment.
Note: If the unit has not been used for an extended
period of time, it needs to operate for several
minutes before power failure alarm can become
activated.
8 - ENG
PN MN239-C4 A | User Manual
Note: The concentrator releases warm air out
the bottom of the unit which can permanently
discolor temperature sensitive flooring surfaces
such as vinyl. The concentrator should not be used
over flooring that is sensitive to heat staining. The
Manufacturer is not responsible for flooring that
becomes discolored.
Note: The NewLife Intensity Oxygen Concentrator
must be operated for at least five minutes at 2 LPM
before using the unit.
The NewLife Intensity is appropriate for usage by
two users, provided the combined flow is a mini-
mum of 2 LPM and does not exceed the maximum
capacity of the concentrator.
Note: The standard NewLife Elite Oxygen Concen-
trator accommodates prescriptions from 1 LPM
minimum to 5 LPM maximum.
Note: The standard NewLife Intensity 8 Oxygen
Concentrator accommodates prescriptions from 2
LPM to 8 LPM maximum.
Note: The standard NewLife Intensity 10 Oxygen
Concentrator accommodates prescriptions from 2
LPM to 10 LPM maximum.
Note: To prevent a void warranty, follow all manu-
facturers' instructions.
Note: Do not attempt any maintenance other than
the possible solutions listed within the manual.
Note: Do not operate the unit without the intake
gross particle filter in place.
Note: Portable and mobile radio frequency (RF)
communications equipment can effect medical
electrical equipment.
Note: There is never a danger of depleting the
oxygen in a room when you use your Oxygen
Concentrator unit.