Interpretation of the results:
DAR
< 1.25
< 1.6
> 1.6
4.1.2. PRIMARY FUNCTION DURING THE MEASUREMENT
During the measurement, the first function of the MODE key is to select the Range of the current:
automatic (default) or fixed.
Display range
Number
Press the MODE key, then the
The choice of Range of current is validated by pressing the MODE key again. The choice remains active
until the switch is turned.
In the Adj. Volt. position, the MODE key can be used to modify the voltage value during the measurement.
4.1.3. SECONDARY FUNCTION
The PRINT secondary function is described in § 6.3 (Printing of measured values).
4.2. DISPLAY / GRAPH KEY
4.2.1. PRIMARY FUNCTION DISPLAY
This key is used to browse through the various accessible screens containing all information available
before, during or after the measurement.
The screens vary depending on the mode selected before the measurement is started.
PI
Condition of insulation
< 1
< 2
< 4
> 4
< 10 M
> 10 M
3
2
key to select the Range, then the
75
Poor or even
dangerous
Good
Excellent
G
1
or
key to modify it.
T
1