Application Programs

All application programs can also be used in legal metrology. Calculated weight
values are indicated by one of the following symbols displayed to the right of the
numerical value:
– Percentage
– Piece count
– Other calculated value
The application programs for calibrated balances not intended for use in legal metrology
are labeled with a triangle G in the right side of the display (e.g. averaging).

Net-Total Formulation/Second Tare Memory

With this application program you can weigh in components for formulation of
a mixture.
Preparation
Configure the Net-Total Formulation/Second Tare Memory application in the operating
menu: See "Configuring the Balance." Menu code: 2.1.3
Example
Step
1. Place an empty container on
the balance
2. Tare the balance
3. Add the first component
4. Store the first component weight.
If the print format is set to include
data ID codes, the following
is printed
5. Add the next component
6. Store the 2nd component weight
7. Add further components,
if desired
8. Display total weight and
fill to desired final weight
PK Elektronik Vertriebs GmbH, E-Mail: info@pkelektronik.com, Internet: www.pkelektronik.com
= %
= pcs
= o
Key (or instruction)
t
v
v
As described
for steps 5 and 6
c
Display/Printout
+ 65.0
g
+
0.0
g
+ 120,5
g
0.0
g
NET
N1
+
120.5 g
+
70,5
g
0.0
g
NET
+ 191,0
g
41