View Window Appearance - BIONIME GM550 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Performing a Test in Order
PRECAUTION
- The glucose test results of blood samples taken from different sites might be varied in certain conditions such
as your glucose is changing rapidly following a drink, a meal, an insulin dose or exercise. In these cases, only
fingertip should be used.
- DO NOT test on the palm or forearm if you are testing for hypoglycemia (Low blood glucose).
- Fingertip samples can show the rapid change of glucose faster than palm or forearm samples.
- Consult with your healthcare professionals whenever you want to conduct a sampler from palm or forearm.
- As the blood flow taken from forearm is slower than fingertip or palm, we recommend using a special lancing
device with Clear Cap for testing sites other than finger.
- If you use normal lancing device for palm or forearm the blood sample might be not enough for meter.
- Please periodically compare the test system to a laboratory test system known to be well maintained and
monitored by a healthcare provider.
32
Make sure your blood sample covers the whole area of the
View Window to get an accurate test result. An insufficient
blood sample will result in an error message ("Er4"). If this
occurs, repeat the test with a new test strip.
PRECAUTION
- Check the expiration date printed on the package every time you use the strip. Do not use expired Rightest
Blood Glucose Test Strip GS550.
®
- Use each Rightest
Blood Glucose Test Strip GS550 immediately after taking it out from the vial.
®
- Do not reuse Rightest
Blood Glucose Test Strip GS550.
- Do the test at least 30 minutes after while entering another site with different ambient temperature.

View Window Appearance

Insufficient blood sample
Enough blood sample
®
33

Publicidad

Capítulos

Tabla de contenido
loading

Productos relacionados para BIONIME GM550

Tabla de contenido