Quality Control Test
®
Please use normal or high level
Rightest
Control Solutions tested with
GM550 under Control Solution Mode. If the test result is within the Control Solution Range printed on the strip vial label,
®
the
Rightest
Blood Glucose Monitoring System GM550 passes Quality Control Test. That means your
Glucose Monitoring System GM550 is working correctly.
Control Solution Range:
Test Strips
Example of Control Solution Range
printed on your test strip vial label.
When should a Quality Control Test be performed ?
®
- To ensure that your
Rightest
Blood Glucose Meter GM550 and Test Strip GS550 are working together properly.
- To confirm that you are following the correct procedure.
- To prepare for your initial blood glucose test.
®
- To check the
Rightest
Blood Glucose Test Strip GS550 when you open a new vial of strips.
®
- To check your
Rightest
Blood Glucose Meter GM550 after it has been dropped, damaged or exposed to liquids.
- When you suspect that your test results are inaccurate, or if your test results are not consistent with the way you feel.
36
®
Rightest
Blood Glucose Monitoring System
Rightest
Control Solution Range mg/dL
Normal:
83-113
High:
237-321
NOTE
- If you want to purchase new normal or high level
®
Blood
Bionime representative.
The possible reasons your Control Solution results are out of the range :
®
- Your Rightest Control Solution has expired or has been opened 3 months ago.
®
- Your Rightest
Blood Glucose Test Strip GS550 has expired.
- You left the cap of the Rightest
- You didn't perform the test procedure correctly.
- Malfunction of the Rightest
If Control Solution results are out of range, your Rightest
Repeat the Quality Control Test. If your Control Solution results outside the range still exist, do not use the Blood Glucose Monitoring
System GM550 to test your blood glucose and contact your Bionime Customer service.
PRECAUTION
Each time you open a new bottle of
the discard date on the label.
months after opening the bottle, or until the expiration date printed on
the label, whichever comes first.
Rightest ®
Control Solutions, please contact your authorized
®
Blood Glucose Test Strip GS550 vial opened or the Rightest
®
Blood Glucose Meter GM550 or Test Strip GS550.
®
Blood Glucose Monitoring System GM550 may not be working properly.
Example of
®
Rightest
Control Solution, write
Expiration Date:
®
Rightest
Control Solution is good for 3
Quality Control Test
®
Control Solution off for a long time.
Exp. Date 2014-06