Performing a Quality Control Test
®
1. Take one
Rightest
Blood Glucose Test Strip GS550 from vial and re-cap the vial cap
immediately.
®
2. Insert the
Rightest
Blood Glucose Test Strip GS550 with view window, facing up, into test
strip port.
3. While the blood drop symbol flashing, press and hold the main button for over 3 seconds until
the "
" symbol appears.
4. You will see blinking "
" symbol and "
" symbol symbol on the screen, prompting you to
®
apply
Rightest
Control Solution.
®
5. Shake the bottle of
Rightest
Control Solution well before opening the cap. Then open the
bottle and put the cap on the table.
®
6. Drip a drop of
Rightest
Control Solution on the top of the cap.
7. Gently touch sample entry of the strip with the
Rightest
cap.
8. When you hear a beep (if volume is turned on ) please wait for the test result. The screen will
display the countdown number from 5.
38
1
4
6
®
Control Solution on the top of the
2
9. Tightly replace the cap on the
10. The Control Solution result appears. Compare your Quality Control Test result to the Control Solution
Range printed on the
5
PRECAUTION
OPEN
- Your Control Solution results will not be calculated for average reading but still can be recalled. The Control
Solution Test result will be shown with "
- Our suggestion environment for Control Solution Test is 59-104℉(15-40℃).
7
- Before "
" and "
because the
Rightest
"
" and "
- Don't drip the
Rightest
might be sucked into the bottle of
degeneration of
strip port as well.
- Don't touch the tip of the
Performing a Quality Control Test
®
Rightest
Control Solution bottle.
®
Rightest
Blood Glucose Test Strip GS550 vial label.
" symbol on the screen.
®
" appears, please don't touch the
Rightest
Control Solution to the sample entry on strip
®
Blood Glucose Meter GM550 is still in an internal check. If you do so, the meter will show
". Accompanied by beeps (if volume is turned on).
®
Control Solution to sample entry of the strip directly. The reagent on strip
®
Rightest
Control Solution and might cause the
®
Rightest
Control Solution. Doing this might contaminate the meter via the test
®
Rightest
Control Solution bottle. If you have touched it, please clean up with water.
9
CLOSE